Nagham Shakir.Turkie; Zahraa Naeem Abd Oun
Abstract
The reaction is based on the Lincomycin that has been responded. In an acidic solution, combine HCl and phosphomolybdic acid to produce a white color precipitate that may be measured ...
Read More
The reaction is based on the Lincomycin that has been responded. In an acidic solution, combine HCl and phosphomolybdic acid to produce a white color precipitate that may be measured using an ISNAG-Fluorimeter (homemade instrument). Some chemical and physical variables have been evaluated. The correlation coefficient was r=0.9983. The calibration curve was linear (the range was 0.0051 mmol/L). The minimum detection limit concentration in the linear dynamic range of the calibration curve was 19.935 ng. The RSD percent of lincomycin HCl at 0.3 and 0.8 mmol/L was less than 0.25% (n=8). The approach was used to determine lincomycin HCl in three pharmaceutical tablets. Comparisons were made using a standard expansion approach and a matched t-test between the recently created method and the traditional strategy (UV-Vis spectrophotometry and turbidimetry with a turbidimeter). At the 95% confidence level, there was no significant difference between the two techniques.